Abstract

The aim was to assess the reliability of bulky DNA adducts measurement by means of the 32P-post-labelling assay. The research design consisted of an intramethod reliability study. Buffy coats from 41 subjects were used to obtain two aliquots of 1–5 μg DNA for each subject; bulky DNA adducts were measured using the nuclease P1 32P-post-labelling technique. The reliability of the measurement was assessed by means of the intraclass correlation coefficient (ICC), the distribution of the differences between the two measurements and the limits of agreement. The estimated ICC was 0.977, with a 95% confidence interval between 0.921 and 0.977. The limits of agreement were ±0.44 (DNA adducts per 108 nucleotides). Only three subjects had differences lying out of such limits. Bulky DNA adduct levels measured by the 32P-post-labelling technique showed good reliability. Only one measurement is needed to use DNA adducts as a biomarker of exposure and, possibly, cancer risk. Besides, as a validation analysis, 32P-post-labelling measurements can be repeated in only 20–30% of samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.