Abstract

A new measurement device was constructed to measure the maximal isometric neck muscle forces. In this study the repeatability of the measurements was evaluated in 34 healthy reserve flying conscripts from Finnish Defence Forces. The measurements of the maximal isometric forces were done in four directions: flexion, extension, and right and left lateral flexions. The intraobserver and intraobserver reliability coefficients were calculated for repeated measurements. The system achieved satisfactory repeatability of lateral flexion and extension force measurements. The total variation was small in lateral flexions and extension, and for these directions was only to a small extent dependent on the examiner and the time of measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.