Abstract

Abstract A novel relaxed eddy accumulation (REA) system for aerosol particle flux measurement has been developed and tested. The system consisted of a fast-response sonic anemometer, a flow system, and software for operating the valves and the concentration analysis system. The prototype was used during September–October 2001 at the SMEAR II station of the University of Helsinki. The REA system was operated with a varying threshold for valve switching determined by the running mean standard deviation of the vertical wind speed. Such a varying threshold made the flux proportionality coefficient β independent of observation conditions. Using temperature as a tracer, β was determined to be 0.392 ± 0.002. The system was validated by comparing the carbon dioxide fluxes estimated by REA with the ones measured by the eddy covariance technique. The system was used subsequently for flux measurements of 50-nm aerosol particles and deposition velocity estimation. Observed deposition velocities over a pine forest dur...