Abstract

BackgroundRecently, surgical navigation systems have been widely used to improve the results of various orthopaedic surgeries. However, surgical navigation has not been successful in anterior cruciate ligament reconstruction, owing to its inaccuracy and inconvenience. This study investigated the registration of preoperative and intraoperative data, which are the key components in improving accuracy of the navigation system. MethodsAn accurate registration method was proposed using new optical tracking markers and landmark retake. A surgical planning and navigation system for anterior cruciate ligament reconstruction was developed and implemented. The accuracy of the proposed system has been evaluated using phantoms and eight cadaveric knees. The present study investigated only the registration accuracy excluding the errors of optical tracking hardware and surgeon. ResultsThe target registration errors of femoral tunnelling for anterior cruciate ligament reconstruction in phantoms were found to be 0.24±0.03mm and 0.19±0.10° for the tunnel entry position and tunnel direction, respectively. The target registration errors measured using cadavers were 0.9mm and 1.94°, respectively. ConclusionsThe preclinical experimental results showed that the proposed methods enhanced the registration accuracy of the developed system. As the system becomes more accurate, surgeons could more precisely position and orient the femoral and tibial tunnels to their original anatomical locations.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call