Abstract

This study is the first regional investigation in Najran, Saudi Arabia aimed at investigating radiation dose and image quality of computed tomography (CT) examinations. The survey data was collected from five scanners in four hospitals. For all CT scanners, a correction factor was calculated to measure the weighted computed tomography dose index (CTDIw) using standard dosimetry phantoms. The CTDIw were reported in this study and compared with other countries. It was found that most CTDIw values were close to the European reference levels and in line with the results of similar surveys in the other parts of world. Concerning image quality, 80 % of the scanners were found to be in compliance with the relative international guidelines for all the examined parameters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.