Abstract

The purpose of this study was to evaluate the ex vivo effects of the Coapsys device upon functional mitral regurgitation (MR) in human hearts. We used seven excised hearts from patients who underwent cardiac transplantation. All patients had functional MR of grade 2 or greater associated with dilated (n = 3) or ischemic (n = 4) cardiomyopathy. After the aortic valve was removed, the left ventricle was pressurized from the aorta with saline at a constant pressure. The degree of MR was then subjectively graded from the opened left atrium (from 0 to 4). The last three studies included volumetric measurements of MR. By tightening the device, the mean MR grade was reduced from 3.3 +/- 0.8 to 1.1 +/- 0.4 (p = 0.0002). In the quantitative analysis, mean regurgitation volume was reduced from 1,108 +/- 1,134 ml/min to 236 +/- 89 ml/min (p = not significant). The mitral annular septal-lateral dimension decreased from 2.0 +/- 0.3 cm to 1.6 +/- 0.5 (p = 0.043). The Coapsys device reduced functional MR in the ex vivo study using excised dilated hearts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.