Abstract
The X-ray pattern of a mass of very fine non-distinguishable anatomical structures alters completely from radiograph to radiograph due to the unavoidable movements of the patient during the exposure. The corresponding image component shows noise-like behaviour and is therefore referred to as the anatomical noise. Reducing this component would enhance the quality of the clinical X-ray image and increase the detectability of radiological signal. We have found that by comparing two X-ray images of the same anatomy acquired under slightly different imaging geometry, it is possible to reduce the anatomical noise in one of the image pair. The proposed method, which allows this, is based on the appropriate attenuation in the wavelet domain. The values of attenuating factors for the wavelet coefficients are proportional to the correlation between the corresponding features of both images. This method was tested for different changes in the imaging geometry. In the case of no geometrical changes, only the quantum and the electronic noise are reduced. An effect of de-noising for the investigated images is obvious.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.