Abstract

ObjectivesTo prospectively compare image quality and apparent diffusion coefficient (ADC) quantification for reduced field-of-view (rFOV)- and multi-shot echo-planar imaging (msEPI)-based diffusion weighted imaging (DWI), using single-shot echo-planar-imaging (ssEPI) DWI as the reference. MethodsUnder IRB approval and after informed consent, msEPI, rFOV, and ssEPI DWI acquisitions were prospectively added to clinical prostate MRI exams at 3.0 T. Image distortion was quantitatively evaluated by root-mean-squared displacement (dr.m.s.). Histogram-based quantitative ADC parameters were compared in a sub-set of patients for proven sites of prostate cancer and matched non-cancerous prostate. Three radiologists also independently evaluated the DWI sequences for subjective image quality and distortion/artifact on a 5-point Likert scale. ResultsTwenty-five patients were included (15 with proven sites of cancer). Average dr.m.s. demonstrated a small but statistically significant reduction in distortion for both rFOV and msEPI relative to ssEPI. Quantitative ADC parameters for prostate tumors demonstrated no significant difference across the 3 DWI acquisitions and each acquisition demonstrated a statistically significant decrease in mean ADC for tumor compared to normal prostate. Qualitative reader assessment demonstrated favorable image quality for rFOV and msEPI, more notable for msEPI. ConclusionsrFOV and msEPI DWI techniques achieved reduction in image distortion, improvement in image quality, and maintained reproducible ADC quantification compared to the standard ssEPI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call