Abstract

IntroductionThe Low-profile Visualized Intraluminal Support (LVIS) device is a new generation of self-expanding braided stent recently introduced in China for stent assisted coiling of intracranial aneurysms. The aim of our study is to evaluate the feasibility, safety, and efficacy of the LVIS device in reconstructive treatment of vertebral artery dissecting aneurysms (VADAs).MethodsWe retrospectively reviewed the neurointerventional database of our institution from June 2014 to May 2016. Patients who underwent endovascular treatment of VADAs with LVIS stents were included in this study. Clinical presentation, aneurysmal characteristics, technical feasibility, procedural complications, and angiographic and clinical follow-up results were evaluated.Results38 patients with VADAs who underwent treatment with LVIS stent were identified, including 3 ruptured VADAs. All VADAs were successfully treated with reconstructive techniques including the stent-assisted coiling (n = 34) and stenting only (n = 4). Post-procedural complications developed in 3 patients (7.9%) including two small brainstem infarctions and one delayed thromboembolic event. Complications resulted in one case of minor permanent morbidity (2.6%). There was no procedure-related mortality. The follow-up angiogram was available in 30 patients at an average of 8.3 months (range, 2 to 30 months), which revealed complete occlusion in 23 patients (76.7%), residual neck in five patients (16.7%), and residual sac in two patients (6.7%). The follow-up of 25 aneurysms with incomplete immediate occlusion revealed 22 aneurysms (88%) with improvement in the Raymond class. One aneurysm (3.3%) showed recanalization and required retreatment. Clinical followed-up at 5–28 months (mean 14.1 months) was achieved in 36 patients because two patients died of pancreatic cancer and basal ganglia hemorrhage, respectively. No new neurologic deterioration or aneurysm (re)bleeding was observed.ConclusionsOur preliminary experience with reconstruction of VADAs with the LVIS device demonstrates that this treatment approach is feasible with good short-term angiographic and clinical outcomes. Long-term and larger cohort studies are necessary to determine long-term outcomes of this therapy.

Highlights

  • The Low-profile Visualized Intraluminal Support (LVIS) device is a new generation of selfexpanding braided stent recently introduced in China for stent assisted coiling of intracranial aneurysms

  • All vertebral artery dissecting aneurysms (VADAs) were successfully treated with reconstructive techniques including the stent-assisted coiling (n = 34) and stenting only (n = 4)

  • Our preliminary experience with reconstruction of VADAs with the LVIS device demonstrates that this treatment approach is feasible with good short-term angiographic and clinical outcomes

Read more

Summary

Objectives

The aim of our study is to evaluate the feasibility, safety, and efficacy of the LVIS device in reconstructive treatment of vertebral artery dissecting aneurysms (VADAs). We aimed to report the preliminary experience of the LVIS stent in treating VADAs

Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.