Abstract

A study is presented on the influence of the instrumental time constant value in the determination of fluctuational sensitivity in atomic-absorption flame photometry. Fluctuational concentration limit (detection limit) has been defined as a function of the noise level of the blank reading (blank adjusted at the 0% absorption level), while the signal noise is a direct function of the time constant of the readout system employed. Employing a Beckman 979 atomic-absorption system and recorder, a scale expander unit was modified to provide a time constant control. The behavior of several elements is presented as an example of the dependence of the fluctuational concentration limit on the system time constant. These results, clearly indicating a basic relationship, lead to the establishment and presentation of a new criterion in an attempt to define and to compare concentration limits of a fluctuational nature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.