Abstract
Monte Carlo simulations of lung equivalent materials often involve the density being artificially lowered rather than a true lung tissue (or equivalent plastic) and air composition being simulated. This study used atomic composition analysis to test the suitability of this method. Atomic composition analysis was also used to test the suitability of 3D printing PLA or ABS with air to simulate lung tissue. It was found that there was minimal atomic composition difference when using an artificially lowered density, with a 0.8 % difference in Nitrogen the largest observed. Therefore, excluding infill pattern effects, lowering the density of the lung tissue (or plastic) in simulations should be sufficiently accurate to simulate an inhaled lung, without the need to explicitly include the air component. The average electron density of 3D printed PLA and air, and ABS and air were just 0.3 % and 1.3 % different to inhaled lung, confirming their adequacy for MV photon dosimetry. However large average atomic number differences (5.6 % and 20.4 % respectively) mean that they are unlikely to be suitable for kV photon dosimetry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.