Abstract

Performance data are presented for several elements determined by atomic-absorption flame photometry with the use of a high sensitivity nitrous oxide—acetylene burner. Utilizing the specific advantages of this flame, the effect of gas mixtures, feeding rate, and the height of reaction zone on the sensitivities achieved are presented. Flame profiles of several elements studied will be discussed with emphasis on using multiple-pass optics for increased sensitivities. Information will be presented on the problems of this type of flame. Results are presented in terms of the precision and accuracy of routine determinations under selected optimum system conditions and will be discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.