Abstract

Real-time spatial compound imaging (SonoCT) is an ultrasound technique that uses electronic beam steering of a transducer array to rapidly acquire several (three to nine) overlapping scans of an object from different view angles. These single-angle scans are averaged to form a multiangle compound image that is updated in real time with each subsequent scan. Compound imaging shows improved image quality compared with conventional ultrasound, primarily because of reduction of speckle, clutter, and other acoustic artifacts. Early clinical experience suggests that real-time spatial compound imaging can provide improved contrast resolution and tissue differentiation that is beneficial for imaging the breast, peripheral blood vessels, and musculoskeletal injuries. Future development of real-time spatial compound imaging will help address the bulk of general imaging applications by extending this technology to curved array transducers, tissue harmonics, panoramic imaging, and three-dimensional sonography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.