This study examines the usability and comfort of a behind-the-ear seizure detection device called brain seizure detection (BrainSD) that captures ictal electroencephalogram (EEG) data using four scalp electrodes. This is a feasibility study. Thirty-two patients admitted to a level 4 Epilepsy Monitoring Unit were enrolled. The subjects wore BrainSD and the standard 21-channel video-EEG simultaneously. Epileptologists analyzed the EEG signals collected by BrainSD and validated it using video-EEG data to confirm its accuracy. A poststudy survey was completed by each participant to evaluate the comfort and usability of the device. In addition, a focus group of UT Southwestern epileptologists was held to discuss the features they would like to see in a home EEG-based seizure detection device such as BrainSD. In total, BrainSD captured 11 of the 14 seizures that occurred while the device was being worn. All 11 seizures captured on BrainSD had focal onset, with three becoming bilateral tonic-clonic and one seizure being of subclinical status. The device was worn for an average of 41 hours. The poststudy survey showed that most users found the device comfortable, easy-to-use, and stated they would be interested in using BrainSD. Epileptologists in the focus group expressed a similar interest in BrainSD. Brain seizure detection is able to detect EEG signals using four behind-the-ear electrodes. Its comfort, ease-of-use, and ability to detect numerous types of seizures make BrainSD an acceptable at-home EEG detection device from both the patient and provider perspective.