Abstract

A real time neutron radiography system has been developed at the University of Michigan Phoenix Memorial Laboratory (PML) and has recently been used to test the imaging capabilities of a neutron imaging device developed by Lixi, Inc. of Downers Grove, Ill. This device uses an input phosphor that is high in gadolinium to generate a light image which is then sent through an intensifier stage to provide images that can be viewed by eye, video camera, or standard 35 mm camera. It was determined that this device provides images of much higher resolution and sensitivity than those obtained with the imaging system currently being used at PML. Using computerized image enhancement techniques, the images obtained with the Lixi neutron imaging device can then be futher enhanced or processed to obtain quantitative information on the object being imaged.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.