Abstract
The use of real-time measurement techniques for objectifying occupational exposure situations is rapidly expanding. However, in many cases, the calibration data provided by manufacturers must be completed by aerosols representative of those present in occupational environments. Present on the market for about ten years, the DiSCmini device is one of the few devices that measures personal exposure to nanometer sized aerosols in real time. Laboratory work conducted on 13 devices provided knowledge about their metrological performance. The experimental data, established for a large range of test aerosols, indicate that this instrument tends to overestimate concentration by 30 to 100 %, and underestimate particle diameter by 20 to 30 %.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.