Abstract

We present real-time detection of airborne Vaccinia viruses using quartz crystal microbalance (QCM) in an integrated manner. Vaccinia viruses were aerosolized and neutralized using an electrospray aerosol generator, transported into the QCM chamber, and captured by a QCM crystal. The capture of the viruses on the QCM crystal resulted in frequency shifts proportional to the number of viruses. The capture rate varied linearly with the concentration of initial virus suspensions (8.5x10(8)-8.5x10(10) particlesml) at flow rates of 2.0 and 1.1 lmin. This work demonstrates the general potential of mass sensitive detection of nanoscale biological entities in air.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.