Abstract
There is considerable difference in published molar absorption coefficients for chlorophyll a. This paper involves estimating the molar absorption coefficient in acetone by calibrating against measured Mg using AAS. Chlorophyll can then be routinely estimated using peak areas in diode array detector HPLC. In order to account for the main errors in the estimates, designed linear calibration experiments are required, involving 14 measurements at five levels, with four instrumental and five dilution replicates. Three errors namely, lack-of-fit, dilution and instrumental errors can be calculated for each type of measurement, guiding the experimenter as to where problems may arise in the experimental procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.