Abstract
Magnetic resonance imaging (MRI) generates potent electromagnetic forces in the form of a static, gradient, or pulsed radiofrequency magnetic field that can result in pacemaker malfunction. This report documents a case of rapid cardiac pacing during MRI in a patient with a dual chamber pacemaker. Although the mechanism of rapid cardiac pacing is unclear, it was directly related to radiofrequency pulsing. We postulated that the lead acts as an antenna for radiofrequency energy that interacts with the pacemaker's output circuit, thus, causing cardiac pacing at a cycle length representing a multiple of the repetition time; or perhaps rapid pacing is related to induced currents generated between the MRI unit and the pacing lead.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.