Abstract

A rapid, simple and sensitive reversed-phase high-performance liquid chromatographic (HPLC) method has been developed for the measurement of acyclovir concentrations in human plasma and its use in bioavailability studies is evaluated. Unchanged acyclovir has been quantified without the introduction of an internal standard using the present method. Human plasma proteins were selectively precipitated by the addition of 7% perchloric acid to spiked plasma samples or to the plasma samples obtained after acyclovir administration to human volunteers and the mixture was spun at 1000 g for 10 min. The supernatant was directly injected into a Novaflex C 18 column and detected at 254 nm. The mobile phase consisted of octane sulfonic acid buffer (pH 2.5) and methanol (92:08). The limit of quantitation for acyclovir in plasma was 20 ng/ml, which enabled the determination of the area under the curve (AUC) more precisely, that is, it is much closer to its extrapolated value. The present method has been successfully applied to samples from bioavailability studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.