Abstract
A simple, rapid, sensitive and specific liquid chromatography-tandem mass spectrometry method was developed and validated for quantification of tamsulosin (I), a highly selective alpha1-adrenoceptor antagonist used for the treatment of patients with symptomatic benign prostatic hyperplasia. The analyte and internal standard, mosapride (II) were extracted by liquid-liquid extraction with diethyl ether-dichloromethane (70:30, v/v) using a Glas-Col Multi-Pulse Vortexer. The chromatographic separation was performed on a reverse phase Waters symmetry C18 column with a mobile phase of 0.03% formic acid-acetonitrile (30:70, v/v). The protonated analyte was quantitated in positive ionization by multiple reaction monitoring with a mass spectrometer. The mass transitions m/z 409.1 solidus in circle 228.1 and m/z 422.3 solidus in circle 198.3 were used to measure I and II, respectively. The assay exhibited a linear dynamic range of 0.1-50.0 ng/mL for tamsulosin in human plasma. The lower limit of quantitation was 100 pg/mL with a relative standard deviation of less than 10%. Acceptable precision and accuracy were obtained for concentrations over the standard curve ranges. A run time of 2.0 min for each sample made it possible to analyze a throughput of more than 400 human plasma samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic, bioavailability or bioequivalence studies.