Abstract
A rapid and sensitive reversed-phase high-performance liquid chromatographic method for analysis of fexofenadine in human plasma has been developed and optimized. The analytes were extracted from biological samples by solid-phase extraction on hydrophilic–lipophilic balance cartridges. LC separation was performed on a C18 analytical column (125 mm × 4 mm i.d., 5-μm particles) with 42:58 (v/v) acetonitrile–water adjusted to pH 2.7 with 85% orthophosphoric acid as mobile phase. Fluorescence detection was performed with excitation at 230 nm and emission at 290 nm. The total time for chromatographic separation was 7 min. The method was validated in accordance with EU guidelines by analysis of plasma samples fortified with fexofenadine at concentrations between 0.05 and 800 ng mL−1. Calibration plots were linear in this range. Mean recovery was typically 94.03% and the detection limit was 0.05 ng mL−1. The time required for quantitative analysis is shorter than that required by other methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.