Abstract

An accurate and sensitive reversed-phase high-performance liquid chromatographic method for analysis of sertraline in human serum, using 4-chloro-7-nitrobenzofurazan as pre-column derivatization agent, is described. The drug and an internal standard (azithromycin) were extracted from serum by use of a mixture of diethyl ether and chloroform, and subjected to pre-column derivatization with the reagent. Analysis of the resulting derivatives was performed on a 250 mm × 4.0 mm cyano column with 63:37 (v/v) methanol–sodium phosphate buffer (0.05 M, pH 3.7) containing 2 mL L−1 triethylamine as mobile phase. Detector response was monitored at excitation and emission wavelengths of 470 and 537 nm, respectively. The calibration plot was linear over the concentration range 2–640 ng mL−1. The lower limits of detection and quantification were 0.5 and 2 ng mL−1, respectively. The method was validated for specificity, sensitivity, linearity, precision, accuracy, and stability and shown to be accurate (intra-day and inter-day accuracy from 0.3 to 4.2%) and precise (intra-day and inter-day precision from 2.4 to 15.5%). The drug was detected at concentrations as low as 2 ng mL−1 in 0.5 mL serum and the method described can be easily applied to human single-dose pharmacokinetic studies of sertraline.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.