Abstract

The standard high pressure liquid chromatography (HPLC) method for determining radiochemical purity (RCP) of [ 131I] m-iodobenzylguanidine ([ 131I] mIBG) is cumbersome and time-consuming. A simplified and rapid separation method with the use of a reverse-phase extraction chromatography technique ( i.e. Sep-Pak ® C18 [SPC] cartridge) is described. The new SPC system closely correlates ( r = 0.99) to the standard HPLC method, and the time to complete the SPC analysis is less than 5 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.