Abstract

A high pressure liquid chromatographic (HPLC) method was developed for the assay of prednisone in bulk drug substances and tablets. The sample was dissolved in water-methanol and an aliquot was analyzed by using HPLC. The average recovery of prednisone added to a prednisone tablet composite was 99.5% with a coefficient of variation of 1.07%. Prednisone was determined in 46 tablets (1-50 mg prednisone/tablet) formulated by 22 manufacturers, using the HPLC method and the USP blue tetrazolium assay. The results show that the HPLC method is more specific and faster than the USP method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call