Abstract

A method for activating ISE is proposed that can allow determination of the fluoride concentration at ng mL(-1) level with good precision and accuracy. Fluoride ISE is activated in 0.5 mol L(-1) HClO4 medium and then fluoride is determined in the same medium. The linear range for the determination of fluoride is between 1.00 x 10(-2)-1.00 x 10(-7) mol L(-1), and the detection limit of the method is 1.0 ng mL(-1). The advantage of this method is that it is free from the use of TISAB solution while being, time-saving and labor-saving. A mechanism study of the activation of FISE in HClO4 medium is explained. The method has been used for the determination of trace fluoride in milk and flour with satisfactory results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.