Abstract

A procedure is presented for the rapid evaluation of HPLC stationary phase stability at pH 8.4 or 10.1 using a temperature of 60°C. Mobile phase (MeOH–0.1moll−1 aqueous NaHCO3, 50:50, v/v) is continuously passed through the column with periodic injections of a test solution until the several chromatographic parameters of the resulting chromatograms are degraded. The tests were applied to several commercial and laboratory-made stationary phases. After degradation two of these phases, one commercial and one laboratory-made, were examined by elemental analysis and scanning electron microscopy to elucidate the degradation process.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.