Abstract
We describe an analysis for metformin in plasma and urine by use of cation-exchange "high-performance" liquid chromatography. The assay requires only 0.5 mL of sample and involves pretreatment with trichloroacetic acid containing an internal standard (1-propylbiguanide), followed by centrifugation and injection into the chromatograph. The column eluent is monitored at 230 nm. Metformin and 1-propylbiguanide give retention times of 8 and 10 min, respectively. Within-day and between-day precision (CV) for the assay of plasma containing 2 mg (1.2 X 10(-5) mol) of metformin hydrochloride per liter was 3.6 and 3.9%, respectively. Similar precision was obtained for urine. The limit of detection is about 0.1 mg (6.0 X 10(-7) mol) of metformin hydrochloride per liter. Several commonly used drugs, including other oral hypoglycemic agents, do not interfere with this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.