Abstract

Simulated SV-AUC data for an adeno-associated virus (AAV) sample consisting of four components having closely spaced sedimentation coefficients were used to develop a high-speed protocol that optimized the size distribution analysis resolution. The resulting high speed (45K rpm) SV-AUC (hs-SV-AUC) protocol poses several experimental challenges: 1) the need for rapid data acquisition, 2) increased potential for optical artifacts from steep and fast moving boundaries and 3) the increased potential for convection. To overcome these challenges the protocol uses interference detection at low temperatures and data that are confined to a limited radial-time window. In addition to providing higher resolution AAV SV-AUC data and very short run times (<20 min after temperature equilibration), the need to match the sample and reference solvent composition and meniscus positions is relaxed making interference detection as simple to employ as absorbance detection. Finally, experimental data comparing hs-SV-AUC (at 45K rpm) with standard low-speed (15K rpm) SV-AUC on the same AAV sample demonstrate the size distribution resolution improvement. These experiments also validate the use of a radial-time window and show how quickly data can be acquired using the hs-SV-AUC protocol.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call