Abstract

The first chiral methodology enabling the separation of indacaterol enantiomers was developed in this work by cyclodextrin-electrokinetic chromatography. Indacaterol (IND) is a chiral drug marketed as a pure enantiomer. Then, the separation and quantification of each enantiomer is of great importance for the quality control of pharmaceutical formulations. After selecting the most suitable chiral selector and background electrolyte, two Box-Behnken designs were achieved to optimize the electrophoretic conditions using two different approaches to shorten analysis times: i) decreasing the capillary length, or ii) performing a short-end injection. Indacaterol enantiomers were separated in less than 5 min with a resolution value of 3.6 under the optimal separation conditions: 0.7% (m/v) carboxymethyl-α-cyclodextrin in 50 mM sodium formate buffer (pH 4.0) and using a short-end injection. Then, the analytical characteristics of the method were evaluated and LODs of 0.05 mg/L for S-IND and 0.04 mg/L for R-IND were achieved. Also, the method allowed the detection of a 0.1% enantiomeric impurity (S-IND) in the R-IND-based pharmaceutical formulations. The developed method was applied to the analysis of two pharmaceutical formulations. Percentages of 97 ± 3% and 103 ± 6% of R-IND with respect to the labeled amounts were found.