Abstract

A quick and direct method for measuring tocopherols (α, β+γ and δ) in vegetable oils has been developed using RP-HPLC with UV detection. Previous extraction of tocopherols is not required. The oil is diluted in hexane and an aliquot is mixed with ethanol containing an internal standard (α-tocopherol acetate). The chromatographic system consists of an ODS-2 column with a methanol–water mobile phase. Tocopherols are detected at 292 nm in less than 5 min after injection. The method is precise (RSD=2.69%) and has a high mean recovery (98.14%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.