Abstract
A high-performance liquid chromatographic (HPLC) method for the determination of valsartan in human plasma is reported. The assay is based on protein precipitation with methanol and reversed-phase chromatography with fluorimetric detection. The preparation of a batch of 24 samples takes 20 min. The liquid chromatography was performed on an octadecylsilica column (50 mm x 4 mm, 5 microm particles), the mobile phase consisted of acetonitrile -15 mM dihydrogenpotassium phosphate, pH 2.0 (45:55, v/v). The run time was 2.8 min. The fluorimetric detector was operated at 234/374 nm (excitation/emission wavelength). The limit of quantitation was 98 ng/ml using 0.2 ml of plasma. Within-day and between-day precision expressed by relative standard deviation was less than 5% and inaccuracy did not exceed 8%. The assay was applied to the analysis of samples from a pharmacokinetic study.