Abstract
A rapid, sensitive and selective method for the simultaneous determination of bromadiolone, flocoumafen and brodifacoum in whole blood using warfarin as internal standard (IS) by high-performance liquid chromatography coupled with electrospray ionization mass spectrometry (HPLC/ESI-MS) has been developed and validated. The target compounds were extracted from the whole blood with ethyl acetate and separated on an XDB C18 column (150 mm x 2.1 mm i.d. x 5 microm) by using a mobile phase consisting of 0.2% acetic acid/methanol (12/88, v/v) at a constant flow rate of 0.50 mL/min. The analytes were detected using negative ESI-MS in the selected ion monitoring (SIM) mode. The molecular ions [M-H]- of m/z 527, 541,523 and 307 were selected for the quantification for bromadiolone, flocoumafen, brodifacoum and the IS, respectively. The calibration curves were linear (r2 > 0.995) in the concentration range of 0.50-100.00 ng/mL. The method showed a satisfactory sensitivity (0.05-0.5 ng/mL using 200 microL blood), precision (RSD < 11.9%), accuracy (recovery: 82.0-96.1%) and selectivity. This method was successfully applied to the determination of the analytes for the diagnoses of poisoned human beings and animals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.