Abstract

A sensitive, fast, and accurate method to determine warfarin and coumatetralyl in serum by high performance liquid chromatography coupled with electrospray ionization mass spectrometry (HPLC‐ESI‐MS) was developed. After warfarin and coumatetralyl in serum was extracted by ethyl acetate, a satisfactory separation was achieved on a XDB C18 column, 150 mm×2.1 mm×5 µm, by using the mobile phase consisting of acetic acid‐ammonium acetate (10 mmol/L, pH=4.5) (20%) and methanol (80%) in isocratic eluent at a constant flow rate of 0.30 mL/min. The molecular ions m/z [M‐H]− 307 and 291 were selected for the quantification in the selected ion monitoring (SIM) mode for warfarin and coumatetralyl, respectively. The linear range was 0.5∼100.0 µg/L for warfarin and coumatetralyl. The limit of quantification in serum was found to be 0.5 µg/L for each. The recoveries were between 91.9∼94.9% for warfarin and 90.9∼104.0% for coumatetralyl, the intra‐ and inter‐day RSDs for each were less than 10.1%. This method can determine trace warfarin and coumatetralyl in serum and it can be used to diagnose poisoned humans and other animals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.