Abstract
Fecal lipid content is usually determined by titrimetric or gravimetric methods, but these methods are time consuming and involve dangerous solvents. We have developed a new method of measuring fecal lipids by Fourier transform infrared spectrometry (FTIR) with an attenuated total reflectance accessory that is fast and requires no solvents. The spectra of stools from 4000 to 750 cm-1 were analyzed, and the lipid concentrations were measured by using a calibration curve prepared by partial least-squares analysis of data from 34 stools. The linearity of the method was tested by mixing low-lipid stools with lipid-overloaded stools to give a range of 0.5-15% lipid. The prediction residual values were -0.49-0.78% for calibrators, and -2.55-2.34% for unknown samples. There was good agreement between the fecal lipids measured by gravimetric (x) and FTIR(y) methods: y = 0.87x + 5.5. The standard error of prediction was 1.07%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.