Abstract

A modified carbon rod atomiser was used for the determination of molybdenum in plant material. The sequential introduction of hydrogen, methane, nitrogen and oxygen into the atomiser eliminated the tedious and time-consuming chelation and solvent extraction procedure commonly used for the pre-treatment of samples. Methane also effected continuous re-coating of the atomiser with pyrolytic graphite. An optical device increased the precision of the injection of the samples. The improvement in precision, accuracy, sensitivity, rapidity and repeatability gained through the use of this modification was demonstrated by the determination of molybdenum in certified samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.