Abstract
A reliable and above all, rapid antimicrobial susceptibility test (AST) is required for the diganostics of blood stream infections (BSI). In this study, resistance testing using DxM MicroScan WalkAway (MicroScan) from a 4-h subculture is compared with the standard overnight culture (18-24h). Randomly selected positive blood cultures (PBC, n = 102) with gram-negative bacteria were included in the study. PBC were sub-cultured onto appropriate agar plates and AST by MicroScan was performed after 4 h of incubation and repeated after incubation for 18-24 h as standard. In a total of 1909 drug-strain pairs, the 4-h subculture approach showed a very high essential agreement (EA) (98.6%) and categorical agreement (CA) (97.1%) compared with the standard. The incidence of minor error (mE), major error (ME), very major error (VME), and adjusted very major error (aVME) was 1.1%, 0.4%, 12.9%, and 5.3%, respectively. In summary, the use of 4-h subcultures for resistance testing with the MicroScan offers a very reliable and easy to realize time saving when testing positive blood cultures with gram-negative bacteria.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.