Abstract

A rapid, sensitive, and practical CE with C(4) D detection was developed for the analysis of three polyphenols (rutin, scopoletin, and chlorogenic acid) in tobacco samples. The constructed mini detection cell (12 mm × 10 mm × 10 mm) of C(4) D featured with small inner cell volume (∼2 nL), smaller noise (<0.9 mV), repeatability, high strength and durableness. Three polyphenols were ultrasonically extracted with methanol-water (70:30, v/v) solution following SPE cleanup. The CE method was optimized with the running buffer of 150 mmol L(-1) 2-amino-2-methyl-1-propanol (pH 11.2), and the applied separation voltage of +20 kV over a capillary of 50 μm id × 375 μm od × 50 cm (38 cm to the C(4) D window, 41.5 cm to the UV detector window), which gave a baseline separation of three polyphenols within ca. 6 min. The method provided the limits of quantification (S/N = 10) at about 0.08-0.15 μg g(-1) for three polyphenols, whereas the overall recoveries ranged from 82% to 88%. The proposed method has been successfully applied to measure three polyphenols in the actual tobacco samples, and their contents were calculated and evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.