Abstract

Background In several clinical situations, lung separation and single-lung ventilation (SLV) is essential. In these cases, the double-lumen tube (DLT) is the most widely used device. Bronchial blocker such as Univent or Arndt Blocker serves as an alternative. The EZ-Blocker® (EZ; AnaesthetIQ B.V., Rotterdam, The Netherlands) is a new device promising to exceed clinical performance of DLT. The aim of this study was to assess the clinical performance of EZ in comparison with conventional left-sided DLT. Methods Forty adult patients undergoing elective thoracic surgery requiring thoracotomy and SLV were included in this study. The patients were randomly assigned to one of two groups: EZ (combined with conventional 7.5 or 8.5 mm single-lumen tube) or DLT (37 or 39 Fr left-sided DLT). Time for intubation procedure and time to verification of the correct position of EZ or DLT using fibreoptic bronchoscopy (FOB) were recorded. After surgery, a thoracic surgeon rated the quality of collapse of the lung (1–3 on a three-level scale). Results Time for intubation using DLT 85.5 (54.8) s was significantly faster (P Conclusions Although time for intubation was longer with the EZ, the device proved to be an efficient and easy-to-use device. The EZ is a valuable alternative device to conventional DLT. Verification of the correct position of the EZ by FOB seems to be obligatory. This study was registered at http://www.clinicaltrials.gov (identifier: NCT01171560).