We have demonstrated a sensitivity enhancement factor of 500 in aqueous solutions using a liquid core optical fiber (LCOF) Raman cell made from Teflon-AF. We were able to collect a spectrum of 54 microM lysozyme with a signal-to-noise ratio of 31 in the LCOF Raman cell using 24 mW of laser power and 3 min of integration time. The lysozyme Raman intensity was only 1% of the background Raman intensity from water, but the water-subtracted lysozyme spectrum was still shot-noise-limited and essentially free of nonrandom noise. The lack of nonrandom noise indicates that it should be possible to collect good quality Raman spectra of proteins such as lysozyme at even lower concentrations. The 2.4-microL sample volume of the LCOF Raman cell is an added benefit when limited quantities of sample are available. This volume of a 54 microM lysozyme solution corresponds to only 13 nanomoles or 1.9 microg of lysozyme.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.