Abstract

A microprocessor-controlled instrument for measuring the concentrations of radon progeny in indoor air is described. The measurement technique is based on alpha spectroscopy and uses two counting intervals following a sampling period during which radon progeny are collected on a filter. The counting intervals are selected to provide optimal precision for measuring 222Rn progeny for fixed total measurement times ranging from 30 to 60 min: concentrations as low as 0.5 pCi/1 can be measured with less than 20% uncertainty in 45 min. The instrument can also be used to estimate the potential alpha energy concentration of 220Rn decay products. The device operates under the control of a computer or a data terminal and functions for week-long periods between filter changes. The user can specify the sampling- and counting-interval timing over a wide range and select from among several operating modes. A number of performance tests are also described indicating that for typical indoor concentrations the measurement uncertainty is dominated by counting statistics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.