Abstract
A solid phase radioimmunoassay for gastrin was established by covalent coupling of antibodies to bromoacetylcellulose. An antiserum monodisperse in respect to binding energy was employed. The effect of covalent coupling on antibody-binding energy, antibody utilization, specificity, and assay re-producibility was investigated by comparison with assays employing charcoal, polyethylene-glycol, or ion-exchange resin for separation. Specificity, within and between-assay reproducibility were of the same order for the four methods employed. A decrease in antibody-binding energy, expressed by the equilibrium constant K°, from 1.0 × 1012 l/mol to 0.3 × 1012 l/mol was observed after antibody coupling (mean of four batches). The final dilution of antiserum was 1:37,500 in assays using charcoal, polyethylene-glycol, and ion-exchange resin. The corresponding dilution in the antibody solid phase assay was 1:2000, resulting in a reduction of the number of gastrin determinations per ml antiserum from 15,000 to 800. It is...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian journal of clinical and laboratory investigation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.