Abstract
We describe a fully automated continuous-flow radioimmunoassay for digoxin with use of a Technicon AutoAnalyzer system. It makes use of antibodies linked to magnetizable particles; a magnet separates the bound and free fractions. The precise dispensing of reagents by the AutoAnalyzer system permits a small incubation volume (about 160 microliter) and the reproducible timing enables non-equilibrium conditions to be used. Thus a single sample is assayed in 15 min, with a sample throughput of 30/h. The standard curve ranges from 0.5 to 8.0 microgram/liter and is most precise between 1 and 3 microgram/liter, the between-assay coefficient of variation being less than 3%. There is no significant carryover between samples of high and low concentration, and results by the method correlate closely (r = 0.969) with those by an established manual assay in which charcoal separation is used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.