Abstract
We describe a fully automated continuous-flow radioimmunoassay for methotrexate. [125I]Histamine-labeled methotrexate was used as tracer. Anti-methotrexate serum was coupled to a magnetizable solid-phase and the bound and free fractions were separated with an electromagnetic field. The assay is precise (CV less than 2.5%) and rapid (30 samples per hour), incubation volume is small (about 160 micro L), and incubation brief (10 min). The accurate timing inherent in the system obviates the need to attain equilibrium, so that assay of each sample takes only 15 min. The assay is sensitive (1--100 microgram/L). There is no significant carryover between samples of high and low concentration. Results by the automated method correlated well with those by both a manual assay in which the same reagents and separation technique are used (r - 0.99) and a competitive protein-binding assay (r = 0.96).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.