Abstract
To perform concurrent TMS-fMRI on difficult targets, such as the occipital lobe. a 3-channel flexible, thin RF coil was constructed that allows for whole head image coverage without impeding TMS placement. A custom MR-safe patient table which mates with typical 3T MR scanners was constructed for the purpose of face-down subject positioning to allow for access to posterior TMS targets. To counterbalance the loading effect of the TMS coil on the RF coil circuits "Dummy Loads" are introduced, which mimic the loading conditions of the TMS coil when it is not present. the designed RF coil performed as expected, achieving acceptable SNR values at depths equal to the center of the average human head, and high SNR values near the surface. The system allows for concurrent TMS-fMRI, targeting any area of the being while imaging the entire volume.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.