Abstract

We present an overview of the National Institute of Standards and Technology high temperature blackbodies, both in operation and in development, suitable for heat-flux sensor calibration. Typical results of calibrations using the transfer technique in the 25 mm Variable-Temperature Blackbody are presented to demonstrate the long-term repeatability of the calibration technique. A comparative study of the absolute and transfer calibrations of a Gardon gage in a spherical blackbody with a cooled enclosure surrounding the gage housing was conducted. Results of this study demonstrated the influence of convection associated with absolute calibration of sensors in a cooled enclosure. Plans for further development of the transfer technique to higher heat-flux levels and the associated technical issues are discussed.