Abstract

Radio-chromic film based on polyvinyl butyral (PVB) containing different concentrations of methyl red (MR) dye for 0.125, 0.25 and 0.5 mM has been introduced as high dose dosimeter. The dosimeters were irradiated with gamma ray from 60Co source at doses from 5 to 150 kGy. UV/vis spectrophotometry was used to investigate the optical density of unirradiated and irradiated films in terms of absorbance at 497 nm. The dose sensitivity of MR-PVB film dosimeter increases strongly with increase of absorbed dose as well as increase of concentrations of MR dye. The effects of irradiation temperature, relative humidity, dose rate and the stability of the response of the films after irradiation were investigated and found that these films could be used as routine dosimeter in industrial radiation processing. The useful dose range of developed MR-PVB film dosimeters is in the range of 5–100 kGy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.