Abstract

This study aimed to audit large quantities of mini C-arm data used in hand and wrist surgery and to analyse what effect the type and anatomical location of procedures had on screening time and dose area product. Of a total of 2032 procedures, the median screening time was 11 seconds and median dose area product was 0.75 cGycm2. The third quartile value for screening time was 23 seconds and dose area product was 1.62 cGycm2. The median screening time for closed procedures was 7 seconds and the dose area product was 0.57 cGycm2. The median screening time for open surgery was 23 seconds with a median dose area product of 1.45 cGycm2. The data for steroid injection procedures were similar to those of the closed procedures, with a median screening time of 7 seconds and dose area product of 0.45 cGycm2. We found no significant differences in the screening time or dose area product between procedures on the wrist and forearm compared with the hand and digits. Level of evidence: III.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.