Abstract

Enteroclysis is a minimally invasive radiographic examination of the small intestine. During the procedure, considerable radiation dose is delivered to the patients. This study intends to: (a) evaluate the radiation dose to the patient using thermoluminescent dosemeters (TLDs, according to the protocol used at Radiology Department, University Hospital of Larissa, Greece; (b) estimate the thyroid surface dose (TSD) and doses to some radiosensitive organs located in the irradiation field. A total of 46 patients was examined. Patients were divided into two groups according to the digital X-ray machine used. The mean entrance surface dose (ESD) was 601.2 ± 96.2 mGy and the mean fluoroscopy time was 8.5 ± 3 min, while the mean TSD was 0.34 ± 0.6 mGy. The ESD for group A was lower by 20 % than group B due to fluoroscopic mode used. The dose values were higher than those in the literature. A local diagnostic reference level was introduced for further patient dose optimisation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.