Abstract

Hysterosalpingography (HSG) is an efficient radiological examination for the evaluation of the female reproductive tract. However, it involves unavoidable irradiation to the ovaries of women in childbearing age. Therefore, radiation dose optimisation is required in order to reduce the probability of the associated risks. This study attempts to: measure patient and staff doses, estimate the effective dose and radiation risk for HSG using digital fluoroscopic images. Thirty-seven patients with infertility were examined using two digital X-ray machines. Thermoluminescence dosimeters (TLD) were used to measure entrance surface dose (ESD) for patients and staff during the procedure. The mean ESD and thyroid surface dose of the patient were 3.60 and 0.17 mGy, respectively, while the mean ESD for the staff was 0.18 mGy per procedure. The patient overall risk for cancer and hereditary effects is 24 x 10(-6), while the risk for the staff is negligible. HSG with fluoroscopic technique demonstrate improved dose characteristics, compared to the conventional radiographic-based technique, reducing the surface dose by a factor of 3, without compromising the diagnostic findings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.